Mains connection and power supply
6 - 3
Siemens AG
Register 3
SPB7-230.033.11
Page 3 of 4
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Checking the supply voltages
6
The correct supply vo5 V, +15 V and +24 V have already been tested in the fac-
tory. Only a visual check of the function is therefore needed.
1. System "ON".
2. Check that the following orange diodes are "ON" on p.c. board D704:
V39 (+15 V)
V38 (-15 V)
V41 (+5 V)
V40 (+24 V)