Preparatory
work
3
-
3
Siemens AG
Register 3
SPB7-230.033.11
Page 3 of 8
M1000/3000 Nova
Medical Solutions
Rev. 06
05.06
CS PS SP
Installation and Start-Up
Removing the transport safeguards
3
Removing the swivel-arm system transport safeguard
3
1. Loosen the transport safeguard nuts (1/ Fig. 2) and remove the tube head screws
(2/ Fig. 2).
2. Push the red transport safeguard (3/ Fig. 2) inwards, away from the tube head.
Transport safeguard for the lifting carriage
3
1. Loosen the two screws (1/ Fig. 3) and remove the red transport safeguard (2/ Fig.
3). Reinsert screws (1/ Fig. 3), they might be needed for service purposes.
Fig. 2 Swivel-arm system transport safeguard
Removal of the swivel-arm system transport safeguard is done
after the equipment has been powered up, see further section
"Attaching the swivel-arm covers" on page 7 - 1.
Fig. 3 Transport safeguard for the lifting carriage
M
A
M
0
0
0
8
1
1
2
3
NOTE