GX-6000 Operator’s Manual
Appendix F: ESS-03 Toxic Sensors • 195
Figure 42: Sensor Gasket Seating
14 . Insert the flow chamber back into the instrument.
15 . Tighten the flow chamber’s four screws that were loosened in Step 2.
16 . Start up the GX-6000 by pressing and briefly holding the POWER/ENTER button.
17 . To verify that the sensor gasket was inserted properly:
a. Plug the inlet with your finger.
b. Verify that the GX-6000 goes into low flow alarm. If the GX-6000 does not go
into low flow alarm, turn the instrument off and attempt to seat the sensor gasket
again.
Gasket Aligning Tabs
Flow Fitting Connections