12 • Chapter 1: Introduction
GX-6000 Operator’s Manual
Specifications
Table 1: Standard Sensor Specifications/Alarm Points
Table 2: GX-6000 Specifications
Combustible
Gas, Methane
(CH
4
)
Calibration
Standard**
Oxygen (O
2
)
Hydrogen
Sulfide
(H
2
S)
Carbon
Monoxide
(CO)
VOCs, Isobutylene Calibration
Standard**
Low Range PID
High Range PID
Detection
Range
0 - 100 %LEL
0 - 40
volume%
0 - 100.0
ppm
0 - 500
ppm
0 - 50000 ppb
0 - 6000 ppm
Reading
Increment
1 %LEL
0.1 volume%
0.5 ppm
1 ppm
• 0 - 5000:
1 ppb
• 5000 - 50000:
10 ppb
• 0 - 600 ppm:
0.1 ppm
• 600 - 6000 ppm: 1
ppm
Alarm 1
Factory Setting
10 %LEL*
19.5
volume%,
decreasing*
5.0 ppm*
25 ppm*
5000 ppb*
400 ppm*
Alarm 2
Factory Setting
50 %LEL
23.5
volume%,
increasing
30.0 ppm
50 ppm
10000 ppb
1000 ppm
STEL Alarm
n/a
n/a
5.0 ppm
200 ppm
n/a
60.0 ppm
TWA Alarm
n/a
n/a
1.0 ppm
25 ppm
n/a
40.0 ppm
*
When calibrating the GX-6000 with the Auto Calibration or the Single Calibration method, the calibration gas value must be
equal to or higher than the alarm 1 setting. See “Updating the Alarm Point Settings (ALARM POINTS)” on page 145 for
instructions to change the alarm points if necessary for the desired calibration gas value.
**
Although calibrated to a specific gas, these sensors will respond to a variety of gases.
Sampling
Method
Sample Draw
Response
Time
T90 Within 30 Seconds
Display
Graphics LCD Display
Operating
Temperature
& Humidity
-20°C to 50
°
C/Below 95% RH (Without Condensation)