170 • Appendix C: Using the GX-6000 in Leak Check ModeGX-6000 Operator’s Man
-
Snap Log Mode, Leak Check Mode Operation
The snap logging function in Snap Log Mode allows the user to record data at a specific
time and have it saved to the data logger. The data is assigned a snap log ID and is saved
with the station ID that was selected when the data was taken.
To enter Snap Log Mode and record snap log data:
1.
Turn the GX-6000 on as described in “Turning On the GX-6000, Leak Check Mode”
on page 165.
2 . Select the desired display range. In the example below, 5000 ppm has been selected.
3 . Press and hold the SHIFT
▼
(PANIC) button, then press and hold the
▲
AIR button
and hold both until you hear a beep. The Station ID Select Screen will display and
the current station ID will be in the middle of the screen.
4.
To exit Snap Log Mode and return to Leak Check Mode, press and release the DISP/
LOCK button.
5.
Use the
▲
AIR and SHIFT
▼
(PANIC) buttons to scroll to the desired station ID.
6.
Press and release the POWER/ENTER button to continue to the Base Reading
Screen.
NOTE:
If
CO DISPLAY
is set to
ON
, the CO reading is saved during a snap log but
will not be displayed in the Base Reading Screen, Peak Reading Screen, or
while viewing snap log data at the instrument. However, the saved CO
reading(s) can be viewed using the GX-6000 Datalogging Program.
REC DATA
REC. DATA
STATION ID
TANK 5
NO : DISP
4:07
CH4
F . S .
100
ppm
5000
ppm
BASE