GX-6000 Operator’s Manual
Chapter 3: Operation • 63
Gas Display Screen
The Gas Display Screen allows the user to choose how the gas readings are displayed in
Measuring Mode. A description of the choices can be found in Table 10 below. If anything
other than
DISPLAY ALL
is selected, the instrument will return to
DISPLAY ALL
when
it is turned off and turned back on.
1 . With the Gas Display Screen showing, press and release the POWER/ENTER
button. The available choices will appear.
2 . Use the
▲
AIR and SHIFT
▼
(PANIC) buttons to move the cursor next to the desired
setting.
Table 10: Gas Display Choices
Option
Description
DISPLAY ALL
All active channels’ readings will be displayed at once
SCROLL AUTO
One active channel will display at a time and the instrument will
automatically scroll through all active channels for the duration of the
operating session.
SCROLL MANUAL
One active channel will display at a time and the user must press and
release the POWER/ENTER button to scroll through all active channels.
If the POWER/ENTER button is not pressed and released, the
instrument will display the same channel’s readings for the duration of
the operating session.
DISP
GAS DISP
YES :ENTER
NO :DISP
EXIT :RESET
DISP
GAS DISP
DISPLAY ALL
SCROLL AUTO
SCROLL MANUAL