138 • Appendix A: Maintenance Mode
GX-6000 Operator’s Manual
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the main menu.
Performing a Calibration (GAS CALIBRATION)
1 . From the main menu, place the cursor next to
GAS CALIBRATION
.
2 . Press and release POWER/ENTER. The Gas Calibration Screen appears.
3 . Use the
▲
AIR or SHIFT
▼
(PANIC) buttons to move the cursor to the operation you
wish to perform.
See “Calibrating Using the Auto Calibration Method” on page 71 for instructions to
perform an automatic calibration.
See “Calibrating Using the Single Calibration Method” on page 82 for instructions to
perform a single calibration.
Performing a Bump Test (BUMP TEST)
NOTE:
The
BUMP DISP
menu item in Maintenance Mode is factory set to
OFF
. The
BUMP TEST
menu item will not appear unless
BUMP DISP
is set to
ON
. See
“Updating the Bump Display Setting (BUMP DISP)” on page 141 for
instructions.
Entering the
BUMP TEST
menu item brings you to the first of the Gas Select Screens.
See “Performing a Bump Test” on page 87 for instructions, starting at Step 8.
MAINTE
GAS CALIBRATION
AIR CAL
AUTO CAL
SINGLE CAL
ESCAPE