102 • Chapter 6: Maintenance
GX-6000 Operator’s Manual
Replacing or Recharging the Batteries
WARNING:
To prevent ignition of a hazardous atmosphere, batteries must only be
changed or charged in an area known to be nonhazardous.
AVERTISSEMENT:Pour éviter l’inflammation d’une atmosphère dangereuse, les
batteries doivent uniquement être modifiés ou facturés dans une zone
connue comme non dangereuse.
Replace or recharge the batteries when the battery icon indicates that the unit is in low
battery warning. When in low battery warning, the lowest battery level indication bar
disappears and the battery icon will be flashing.
Replacing Alkaline Batteries
NOTE:
Use Procell PC 1500 alkaline batteries or RKI Instruments, Inc. lithium-ion
battery pack 49-1619 to maintain the CSA classification of the GX-6000. Use
of other batteries will void the CSA classification and may void the warranty.
Do not mix old/new or different types of batteries.
NOTE:
Utiliser Procell 1500 piles alcalines ou RKI Instruments, Inc. pack batterie
lithium-ion 49-1619 de maintenir la classification CSA de la GX-6000.
L’utilisation d’autres piles annule la classification CSA et peut annuler la
garantie. Ne mélangez pas les anciennes/nouvelles ou différents types de piles.
1 . Turn off the GX-6000.
WARNING:
Do not remove the batteries while the GX-6000 is on.
2 . Turn the GX-6000 over so that the flow chamber and battery cover are facing up and
the LCD is facing down.
3 . Push the battery cover latch toward the bottom of the instrument.
Figure 11: Releasing Battery Cover Latch