13 . Press and release the POWER/ENTER button to start a calibration. An
automatically-determined, temperature-based countdown is displayed at the bottom
of the screen. The number to the left of the countdown (“5” in the example below) is
the temperature code.
14 . If the calibration passes, the screen will indicate PASS and will return you to the
TUBE CAL Auto Calibration Menu.
15 . If the calibration fails, a failure screen will appear and the instrument will go into
alarm.
Press and release the RESET button to clear the alarm and return to the TUBE CAL
Auto Calibration Menu. See “Troubleshooting” on page 100.
16 . Use the SHIFT
▼
(PANIC) button to place the cursor next to
ESCAPE
, then press
and release the POWER/ENTER button to return to the TUBE CAL Menu.
APPLYGAS
BNZ
5 45SEC
0.00
ppm
TUBE CAL
BNZ
FAIL
ppm