214 • Appendix H: 10.0 eV/Benzene PID Sensor
GX-6000 Operator’s Manual
11 . To start another measurement, press and release the POWER/ENTER button.
WARNING:
If you take the GX-6000 to an area that is 10 or 15 degrees different
than the previous area, allow the instrument 45 minutes to acclimate
before taking any readings. Failure to do so may cause the reading to
be inaccurate.
a. You will be prompted to change the tube before continuing. See “Replacing a
Tube” on page 229 for instructions.
b. Once the tube is changed, press and release the POWER/ENTER button to start
the measurement. Return to Step 8.
12 . To not start another measurement and to return to Measuring Mode, press and release
the DISP/LOCK button.
a. You will be prompted to remove the tube holder from the instrument’s inlet
fitting.
b. Remove the tube holder from the GX-6000’s inlet fitting.
c. Press and release the POWER/ENTER button.
TUBE
CHANGE TUBE
YES :ENTER
NO :DISP
EXIT :RESET
TUBE
REMOVE HOLDER
YES :ENTER
NO :DISP