GX-6000 Operator’s Manual
Appendix F: ESS-03 Toxic Sensors • 189
7 . Press and release the POWER/ENTER button. The first of the Gas Select Screens,
the standard four sensors screen, will be displayed. In addition to the standard four
sensors screen, there is a separate screen for each installed smart sensor, and an
ESCAPE
screen that will take you back to the Auto Cal menu. Use the
▲
AIR and
SHIFT
▼
(PANIC) buttons to scroll through the different screens.
8 . With the ESS-03 channel’s Gas Select Screen displayed, press and release the
POWER/ENTER button.
9 . If you are calibrating anything other than an HCN sensor, continue to Step 12.
AUTO
GAS SELECT
CH4 50 %LEL
O2 12.0 %
H2S 25.0 ppm
CO 50 ppm
AUTO
GAS SELECT
VOC 100 ppm
AUTO
GAS SELECT
ESCAPE
▲
AIR or
SHIFT
▼
(PANIC)
AUTO
GAS SELECT
HCN 10.0 ppm
▲
AIR or
SHIFT
▼
(PANIC)
▲
AIR or
SHIFT
▼
(PANIC)
▲
AIR or
SHIFT
▼
(PANIC)