GX-6000 Operator’s Manual
Chapter 3: Operation • 65
Combustible Sensor Protection Screen
If combustible gas levels greater than 100 %LEL are expected, the catalytic LEL sensor
can be disabled to protect it. If the catalytic LEL sensor is disabled, there will be no %LEL
alarms in Measuring Mode.
1 . With the
CAT (LEL) SENSOR PROTECTION
screen displayed, press and release
the POWER/ENTER button. The current setting will be displayed along with a
warning about no %LEL alarms if the catalytic LEL sensor is turned off.
2 . Use the
▲
AIR or SHIFT
▼
(PANIC) button to display the desired setting.
NOTE:
If the catalytic LEL sensor is disabled and if the instrument is then turned off
and back on, the catalytic LEL sensor will be enabled again.
3 . Press and release the POWER/ENTER button to return to the
CAT (LEL) SENSOR
PROTECTION
screen in Display Mode.
DISP
CAT (LEL) SENSOR
PROTECTION
DISP
CAT (LEL) SENSOR
***WARNING***
NO LEL ALARMS IF
CAT (LEL) IS
DISABLED
ENABLED