144 • Appendix A: Maintenance Mode
GX-6000 Operator’s Manual
Updating the Bump Expired Setting (BUMP EXPRD)
This item defines what indication is given during start up when a bump test is due and
BUMP REMINDER
is set to
ON
.
With
BUMP EXPRD
set to
CONFIRM TO USE
(factory setting), the GX-6000 will
give an indication at start up if a bump test is past due and will require the user to decide
whether to perform a bump test or continue and use the GX-6000 without bump testing.
Press and release RESET to continue without calibrating or POWER/ENTER to enter
Calibration Mode.
With
BUMP EXPRD
set to
CANNOT USE
, if the unit is due for bump testing, the GX-
6000 will give an indication at start up that a bump test is past due and will prompt you to
press and release POWER/ENTER to enter Calibration Mode and perform a bump test.
Using any other button will have no effect. A successful bump test must be performed in
order to use the instrument.
NOTE:
Even if the password function is turned on in
PASSWORD
, no password will
be required to perform a bump test during startup.
NOTE:
If
BUMP DISP
is set to
OFF
, the bump test menu item will not appear in
Calibration Mode even though the instrument is prompting you to perform a
bump test. A bump test can always be performed in Maintenance Mode, if
necessary.
With
BUMP EXPRD
set to
NO EFFECT
, the GX-6000 will give an indication at startup
that a bump test is past due. If desired, the POWER/ENTER button can be pressed to enter
Calibration Mode and perform a bump test but it is not necessary to acknowledge the
bump test due indication. The warm-up sequence will continue on its own.
1 . From the
BUMP SETTING
screen, place the cursor next to
BUMP EXPRD
.
2 . Press and release POWER/ENTER. The current Bump Expired setting will be
displayed.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument will return to
the
BUMP SETTING
screen.
Updating the Bump Check Gas Setting (BUMP CHECK GAS)
This item defines which sensors are used to update the bump test reminder screen. With
BUMP CHECK GAS
set to
4 GAS
(factory setting), the bump test dates for only the
standard 4 sensors (LEL, O2, H2S, and CO) are used to determine if a bump test is due.
With
BUMP CHECK GAS
set to
ALL GAS
, the bump test dates for the standard 4
sensors along with the PID sensor and any other installed smart sensors are used to
determine if a bump test is due.
1 . From the
BUMP SETTING
screen, place the cursor next to
BUMP CHECK GAS
.
2 . Press and release POWER/ENTER. The current Bump Check Gas setting will be
displayed.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting. The instrument will return to
the
BUMP SETTING
screen.