GX-6000 Operator’s Manual
Appendix H: 10.0 eV/Benzene PID Sensor • 229
4 . The current calibration code will be displayed and it will be flashing.
5 . Use
▲
AIR or SHIFT
▼
(PANIC) to adjust the calibration code to match the one
listed on your filter tube box’s label.
6 . Press and release the POWER/ENTER button to confirm the calibration code. The
screen will show “END” and will return to the Calibration Mode menu.
7 . Use the SHIFT
▼
(PANIC) button to place the cursor next to the
NORMAL MODE
menu option, then press and release the POWER/ENTER button to return to
Measuring Mode.
Maintenance
Replacing a Sensor
See “Replacing a Sensor” on page 113 for sensor replacement instructions.
PID Sensor Maintenance
See “PID Sensor Maintenance” on page 117 for PID sensor maintenance instructions
keeping in mind the following differences in spare parts.
CAUTION:
The aluminum oxide cleaning kit described on pg.117 is only for use with
the 10.0 eV and 10.6 eV PID sensors’ lamps. DO NOT USE THE
ALUMINUM OXIDE CLEANING KIT TO CLEAN THE 11.7 eV PID
SENSOR’S LAMP!
Replacing a Tube
When to Replace Tube
RKI Instruments, Inc. recommends that the tube be replaced before starting a new
measurement in Benzene Select Mode. If discoloration of the filter material extends
beyond the “3/4” line (into the side without the text and arrow), the tube needs to be
replaced. Even though discoloration can be a measure of when to replace the tube, not all
gases cause the filter material to discolor which is why RKI Instruments, Inc. recommends
Part Number
Description
33-0563
PID sensor electrode stack, 10.0 eV/benzene sensor
51-1503
PID sensor replacement lamp, 10.0 eV/benzene sensor
PID-003L
PID sensor, 10.0 eV/benzene
CAL CODE
INPUT
CAL CODE : C