GX-6000 Operator’s Manual
Appendix H: 10.0 eV/Benzene PID Sensor • 223
7 . Press and release the POWER/ENTER button to enter the TUBE CAL Auto
Calibration Menu.
8 . To view or change the benzene concentration to be used for the TUBE CAL:
a. Use the SHIFT
▼
(PANIC) button to place the cursor next to
CONCENTRATION
.
b. Press and release the POWER/ENTER button. The current concentration will be
shown and it will be flashing.
c. Use
▲
AIR or SHIFT
▼
(PANIC) to adjust the concentration.
d. Press and release the POWER/ENTER button to save it.
9 . Screw the demand flow regulator onto the benzene calibration cylinder.
10 . Attach the calibration tubing to the demand flow regulator.
11 . Attach the other end of the calibration tubing to the tube holder’s inlet fitting.
12 . Use the SHIFT
▼
(PANIC) button to place the cursor next to
START CAL
.
TUBE CAL
GAS CALIBRATION
CONCENTRATION
START CAL
ESCAPE
TUBE CAL
CONCENTRATION
BNZ 5.00 ppm