156 • Appendix A: Maintenance Mode
GX-6000 Operator’s Manual
2 . Press and release POWER/ENTER. The Log Over Write Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
LOG
SETTING
screen.
Returning to the Maintenance Mode Menu (ESCAPE)
1 . From the
LOG SETTING
screen, place the cursor next to
ESCAPE
.
2 . Press and release POWER/ENTER. The instrument will return to the main menu.
Adjusting the Low Flow Setpoint (FLOW ADJUST)
If your instrument repeatedly goes into low flow alarm and you have tried all of the
recommendations in the Troubleshooting section, you may need to perform a low flow
setpoint adjustment.
1 . Use the
▲
AIR or SHIFT
▼
(PANIC) button to place the cursor next to
FLOW
ADJUST
.
2 . Press and release the POWER/ENTER button. The pump will turn off and the screen
will display values that reflect the pump’s current draw.
3 . Use the
▲
AIR and SHIFT
▼
(PANIC) buttons to get a reading of approximately
1500 on the bottom value. When you have adjusted the value as close as you can,
press and release the POWER/ENTER button.
4 . The pump will turn on and two new values will be displayed. The top will be a
reference value and the bottom will reflect the pump’s current draw.
MAINTE
FLOW ADJUST
50
1500
MAINTE
FLOW ADJUST
1500
2500