88 • Chapter 4: Calibration Mode
GX-6000 Operator’s Manual
7 . Move the cursor next to the
BUMP TEST
menu item by using the SHIFT
▼
(PANIC) button.
8 . Press and release the POWER/ENTER button. The first of the Gas Select Screens,
the standard four sensors screen, will be displayed. In addition to the standard four
sensors screen, there is a separate screen for each installed smart sensor, and an
ESCAPE
screen that will take you back to the Calibration Mode menu. Use the
▲
AIR and SHIFT
▼
(PANIC) buttons to display the gas(es) you want to test.
9 . Before proceeding, confirm that the bump test gas value(s) are the same as the
concentration(s) in the calibration cylinder. If they are not, adjust the bump test gas
value(s) be entering the
AUTO CAL
menu item in Calibration Mode, changing the
value(s) there, and reentering the
BUMP TEST
menu item.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
BUMP TEST
NORMAL MODE
AUTO
GAS SELECT
CH4 50 %LEL
O2 12.0 %
H2S 25.0 ppm
CO 50 ppm
AUTO
GAS SELECT
VOC 100 ppm
AUTO
GAS SELECT
ESCAPE
▲
AIR or
SHIFT
▼
(PANIC)
▲
AIR or
SHIFT
▼
(PANIC)
▲
AIR or
SHIFT
▼
(PANIC)