162 • Appendix B: Maintenance Mode 2
GX-6000 Operator’s Manual
10 . The main menu displays.
11 . Use the
▲
AIR or SHIFT
▼
(PANIC) button to move the cursor up and down through
the menu items.
Tips for Using Maintenance Mode 2
•
When in the main menu, the cursor (>) indicates which menu item will be selected if
the POWER/ENTER button is pressed and released.
•
Use the SHIFT
▼
(PANIC) button to move the cursor down through the main menu
and submenu items, and to lower values or change the setting in a specific option.
•
Use the
▲
AIR button to move the cursor up through the main menu and submenu
items, and to raise values or change the setting in a specific option.
•
Use the POWER/ENTER button to enter a menu item with the cursor next to it and to
enter and save settings during programming.
•
An adjustable parameter that is flashing can be adjusted with the
▲
AIR and SHIFT
▼
(PANIC) buttons.
•
Press the DISP/LOCK button while in a screen where you are entering or updating
parameters to exit the screen without saving any changes. You can also use the DISP/
LOCK button to back out of submenus and return to the main menu.
Changing the Gas Combination (GAS COMB)
The
GAS COMB
menu item allows you to turn channels on and off.
NOTE:
A smart sensor channel cannot be turned off if the smart sensor is still installed.
If you want to turn off a smart sensor’s channel, you must first remove the smart
sensor and replace it with a dummy sensor.
1 . From the main menu, place the cursor next to
GAS COMB
.
MAINTE 2
MAINTENANCE 2
GAS COMB
HC SELECT
START MEASURE