176 • Appendix D: Using the GX-6000 in Bar Hole Mode GX-6000 Operator’s Manual
Start Up, Bar Hole Mode
This section explains how to start up the GX-6000 in Bar Hole Mode and get it ready for
operation.
Turning On the GX-6000, Bar Hole Mode
WARNING:
Gas alarms are not active when the GX-6000 is in Bar Hole Mode.
CAUTION:
If any sensors other than an IR CH
4
/IR HC or oxygen sensor is installed,
these sensors will not be active while the GX-6000 is in Bar Hole Mode.
NOTE:
In order for
BAR HOLE MODE
to appear as a selection in the Mode Select
Screen, the
L./B. MODE
item in Maintenance Mode must be set to
BAR
HOLE MODE
or
LEAK CHECK MODE BAR HOLE MODE
. The factory
setting is
OFF
.
1 . Connect the tapered rubber nozzle or the sample hose and probe to the GX-6000’s
quick connect inlet fitting.
2 . Press and briefly hold down the POWER/ENTER button. Release the button when
you hear a beep. The Normal Mode Select Screen displays.
3 . Use the
▲
AIR or SHIFT
▼
(PANIC) button to scroll to the Bar Hole Mode Select
Screen.
4:07
MODE SELECT
NORMAL MODE
CH4 100 %LEL
O2 40.0 %
H2S 100.0 ppm
CO 500 ppm
VOC 6000 ppm
4:07
MODE SELECT
BAR HOLE MODE