16 • Chapter 2: Description
GX-6000 Operator’s Manual
Control Buttons
Five control buttons are located below the LCD. They are, from left to right and top to
bottom, DISP/LOCK,
▲
AIR, SHIFT
▼
(PANIC), POWER/ENTER, and RESET.
Flashlight LED
A white LED is visible through a round, raised, frosted lens in the top of the case. This
LED can be used to provide extra light, if necessary. The RESET button turns this LED on
and off.
Alarm LEDs
Five sets of red alarm LEDs (light emitting diodes) border the LCD. The alarm LEDs alert
you to gas, low battery, and failure alarms.
Infrared Communications Port
An infrared (IR) communications port is located on the front of the case, below the
POWER/ENTER button. The data transmitted through the port is in standard IrDA
protocol. A computer’s infrared port or an IrDA/USB cable connected to a USB port can
be used to download data saved by the GX-6000 to a computer using the GX-6000 Data
Logger Management Program. See the GX-6000 Data Logger Management Program
operator’s manual for data logging and downloading instructions.
Buzzer
One solid-state electronic buzzer is located inside the case. Holes on the top front of the
case allow the sound to exit the case. The buzzer sounds for gas alarms, malfunctions, low
battery voltage, and as an indicator during use of the GX-6000’s many display and
Table 3: GX-6000 Control Button Functions
Button
Function(s)
DISP/LOCK
• activates Display Mode
• enters instructions into the GX-6000’s microprocessor
• allows you to exit a menu without saving changes
• locks screen orientation in any mode if
INVERSION SELECT
is set
to
ON
▲
AIR
• activates the demand zero function (adjusts the GX-6000’s fresh air
reading)
• enters instructions into the GX-6000’s microprocessor
• moves the cursor on the LCD up the screen
• increases the value of a parameter available for adjustment
• scrolls through parameter options
SHIFT
▼
(PANIC)
• enters instructions into the GX-6000’s microprocessor
• moves the cursor on the LCD down the screen
• decreases the value of a parameter available for adjustment
• scrolls through parameter options
• initiates a Panic alarm in any mode if
PANIC
is set to
ON
POWER/ENTER
• turns the GX-6000 on and off
• enters instructions, values, and settings into the GX-6000’s
microprocessor
RESET
• silences and resets audible alarm if
ALARM LATCHING
is set to
LATCHING
and
ALARM SILENCE
is set to
ON
• turns the flashlight LED on and off