GX-6000 Operator’s Manual
Appendix H: 10.0 eV/Benzene PID Sensor • 217
Viewing Benzene Select Mode Data
Readings taken in Benzene Select Mode are saved to the snap logger. To view Benzene
Select Mode readings, do the following:
1 . From Measuring Mode, press and release the DISP/LOCK button to enter Display
Mode.
2 . Continue pressing the DISP/LOCK button until the snap logging screen appears.
3 . Press and release the POWER/ENTER button. The screen that appears will depend
on whether or not any data has been saved to the snap logger.
4 . If snap logs have been taken, the screen indicates the year, month, day, and time that
the most recent snap log was taken.
The number near the bottom of the screen indicates the snap log ID number. The first
snap log that is taken is given an ID of 001. The next snap log ID is 002. The ID
number increases sequentially with each set of snap log data.
The last line of the screen indicates the Station ID that was used for the snap log.
5 . Use the
▲
AIR and SHIFT
▼
(PANIC) buttons to scroll through different snap log
IDs.
DISP
REC. DATA DISP
YES :ENTER
NO :DISP
EXIT :RESET
REC DATA
REC. DATA
NO DATA
REC DATA
DATE
5/20/2015
MM/DD/YYYY
3:40
NUMBER 001
TANK 5
No data saved Data saved