160 • Appendix B: Maintenance Mode 2
GX-6000 Operator’s Manual
Appendix B: Maintenance Mode 2
Overview
This appendix describes the GX-6000 in Maintenance Mode 2. The GX-6000 is factory-
set to suit most applications. Update settings in Maintenance Mode 2 only if required for
your specific application. A description of the Maintenance Mode 2 items is shown in
Table 18 below.
Entering Maintenance Mode 2
WARNING:
The GX-6000 is not in operation as a gas detector while in Maintenance
Mode 2.
1 . Take the GX-6000 to a non-hazardous location and turn it off if it is on.
2 . Press and hold the
▲
AIR and SHIFT
▼
(PANIC) buttons, then press and hold the
POWER/ENTER button. When you hear a beep, release the buttons.
3 . The screen that appears will depend on the setting of User Mode’s
PASSWORD
item.
If User Mode’s
PASSWORD
is set to
OFF
(factory setting), continue with Step 4.
If User Mode’s
PASSWORD
is set to
ON
, continue with Step 6.
Table 18: Maintenance Mode 2 Menu Items
Menu Item (Page # of Description)
Description
GAS COMB (pg.162)
Turn channels on or off
HC SELECT (pg.163)
Set the catalytic LEL channel’s detection gas
STREAMING MODE (pg.164)
Turn Streaming Mode on or off
START MEASURE (pg.164)
Enter Measuring Mode
PASSWORD
MODE SELECT
USER MODE
MAINTENANCE
PASSWORD
INPUT
PASSWORD
0 0 0 0
User
PASSWORD OFF
User
PASSWORD ON
User
PASSWORD ON
Maint.
PASSWORD ON
Maint.
PASSWORD
ON
Maint.
PASSWORD OFF
or
OFF
4:07
INPUT
PASSWORD
- - - -
ENTER FOR
MAINTENANCE MODE