GX-6000 Operator’s Manual
Chapter 6: Maintenance • 107
4 . Place the GX-6000 into the battery charging station as shown in Figure 18 below so
that the metal contacts on the back of the unit come into contact with the metal
contacts on the back of the holder in the charging station. When proper contact has
been made, the red LED on the charging station will turn on.
If you are using a DC adapter, the charging station comes with a Velcro strap and
washer assembly to secure the GX-6000. Ensure that the washer is installed over the
GX-6000’s inlet fitting, that both ends of the Velcro strap are attached to the charging
station, and that the assembly is securing the instrument.
Figure 18: Putting the GX-6000 into the Charging Station
5 . When a full charge has been reached, approximately 3 hours, the red LED on the
charging station will turn off. Remove the GX-6000 from the charging station and
unplug the charging station’s power cord from the AC outlet.
Recharging the Lithium Ion Battery Pack Out of the Instrument
The lithium ion battery pack may be charged using the charging station while it is out of
the GX-6000. This is useful if spare battery packs are kept in case the pack in the GX-
6000 needs to be charged, but the unit must be used immediately. In this case, a spare
charged pack can be installed in the GX-6000 and the dead pack charged in the charging
station.
The charger can be mounted to the wall using DIN rail, if desired. See “DIN Rail
Mounting Assembly” on page 22 for instructions.
R E SE T
S H IF T
A IR
( P A N IC )
LOC K
D IS P
RKI GX-6000
Red Charging LED
P O W E R /E N TE R