GX-6000 Operator’s Manual
Appendix A: Maintenance Mode • 145
Returning to the Maintenance Mode Menu (ESCAPE)
1 . From the
BUMP SETTING
screen, place the cursor next to
ESCAPE
.
2 . Press and release POWER/ENTER. The instrument will return to the main menu.
Updating Alarm Parameters (ALARM SETTING)
The
ALARM SETTING
menu item in Maintenance Mode has a sub menu with 4 menu
items:
ALARM POINTS
,
ALARM LATCHING
,
ALARM SILENCE
, and
ESCAPE
.
1 . From the main menu, place the cursor next to
ALARM SETTING
.
2 . Press and release POWER/ENTER. The
ALARM SETTING
menu appears.
Updating the Alarm Point Settings (ALARM POINTS)
This menu item allows you to update one or more alarm points (the reading at which the
GX-6000 recognizes the alarm condition). See Table 1 on page 12 for alarm point factory
settings.
1 . From the
ALARM SETTING
menu, place the cursor next to
ALARM POINTS
.
2 . Press and release POWER/ENTER. The Alarm Points Screen appears and all
detection channels are displayed.
MAINTE
ALARM SETTING
ALARM POINTS
ALARM LATCHING
ALARM SILENCE
ESCAPE
ALARM - P
ALARM POINTS
CH4
O2
H2S
CO
VOC
ESCAPE