134 • Appendix A: Maintenance Mode
GX-6000 Operator’s Manual
Entering Maintenance Mode
WARNING:
The GX-6000 is not in operation as a gas detector while in Maintenance
Mode.
1 . Take the GX-6000 to a non-hazardous location and turn it off if it is on.
2 . Press and hold the
▲
AIR and SHIFT
▼
(PANIC) buttons, then press and hold the
POWER/ENTER button. When you hear a beep, release the buttons.
LOG SETTING
(pg.155)
LOG INTERVAL (pg.155)
• 10 SEC
• 20 SEC
• 30 SEC
• 60 SEC
• 120 SEC
• 180 SEC
• 300 SEC
• 600 SEC
300 SEC
LOG CLEAR DISP (pg.155)
• ON
• OFF
ON
LOG OVER WRITE (pg.155)
• ON
• OFF
ON
ESCAPE
return to main menu
N/A
FLOW ADJUST (pg.156)
set the low flow alarm
N/A
ROM/SUM (pg.157)
view the ROM number and
checksum
N/A
PASSWORD (pg.157)
• ON
• OFF
ON
(0006)
RESTORE DEFAULT (pg.158)
restore default settings
N/A
START MEASURE (pg.159)
enter Measuring Mode
N/A
*
PID channel’s zero suppression is always 0% regardless of what value is shown.
Table 17: Maintenance Mode Menu Items
Menu Item (Page # of Description)
Choices or Action
Factory Setting