84 • Chapter 4: Calibration Mode
GX-6000 Operator’s Manual
4 . Move the cursor next to the
SINGLE CAL
menu item by using the
▲
AIR or SHIFT
▼
(PANIC) button.
5 . Press and release the POWER/ENTER button. The Select Sensor Screen appears.
6 . Move the cursor next to the sensor you want to calibrate with the
▲
AIR and SHIFT
▼
(PANIC) buttons. In the example above, the CH
4
sensor is selected for span
adjustment.
If you do not want to proceed with the span adjustment, press and release the DISP/
LOCK button or place the cursor next to
ESCAPE
and press and release POWER/
ENTER to return to the Calibration Mode Screen.
To proceed with the calibration, continue with the next step.
7 . For a CH
4
, O
2
, H
2
S, or CO channel continue to Step 10.
For a VOC channel, see Step 8 and Step 9 before continuing with Step 10.
CAL
CAL MODE
AIR CAL
AUTO CAL
SINGLE CAL
NORMAL MODE
SINGLE
SINGLE CAL
CH4
O2
H2S
CO
VOC
ESCAPE