174 • Appendix C: Using the GX-6000 in Leak Check ModeGX-6000 Operator’s Man
-
Snap log data that was taken in Leak Check Mode operation will have 2 sequential
screens that are part of one data set. One screen is for the base reading and the other
screen is for the peak reading. “BASE” and “PEAK” will appear in the gas reading
screen for each type of reading, respectively. If a base data point appears without a
peak data point after it, it is because a peak data point was not taken.
NOTE:
Even if
CO DISPLAY
was set to
ON
when the Leak Check snap logs were
taken, the CO reading will not be displayed. The CO readings can be viewed
using the GX-6000 Datalogging Program.
7 . Use the
▲
AIR and SHIFT
▼
(PANIC)
buttons to scroll through the different
snap log data screens. The gas readings will change but the snap log ID
is not visible from this screen.
8 . You can also go back and forth between the ID and data screens by pressing and
releasing the POWER/ENTER button.
9 . To return to Leak Check Mode, press and release the DISP/LOCK button.
Turning Off the GX-6000, Leak Check Mode
1 . Press and hold the POWER/ENTER button.
2 . TURN OFF will appear on the display and the buzzer will pulse for about five
seconds.
3 . Release the button when TURN OFF disappears from the display.