GX-6000 Operator’s Manual
Appendix A: Maintenance Mode • 147
11 . When you are done making changes to all channels, use SHIFT
▼
(PANIC) to move
the cursor next to
ESCAPE.
12 . Press and release POWER/ENTER to return to the
ALARM SETTING
menu.
Updating the Alarm Latching Setting (ALARM LATCHING)
With
ALARM LATCHING
set to
LATCHING
(factory setting), the GX-6000 remains
in alarm condition until the alarm condition passes
and
the RESET button is pressed.
With
ALARM LATCHING
set to
SELF RESET
, the GX-6000 automatically resets an
alarm when the alarm condition passes.
1 . From the
ALARM SETTING
menu, place the cursor next to
ALARM
LATCHING
.
2 . Press and release POWER/ENTER. The Alarm Latching Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
ALARM
SETTING
menu.
Updating the Alarm Silence Setting (ALARM SILENCE)
With
ALARM SILENCE
set to
ON
(factory setting), pressing and releasing the RESET
button silences the buzzer when the GX-6000 is in alarm. The LEDs continue to flash, the
vibrator continues to pulse, and the display continues to show the alarm. If you enter
Display Mode during an alarm condition, the buzzer will be silenced but the LEDs will
continue to flash and the vibrator will continue to pulse. If you return to Measuring Mode
and there is still an alarm condition, the LEDs will continue to flash, the vibrator will
continue to pulse, and the buzzer will remain off.
With
ALARM SILENCE
set to
OFF
, you cannot silence the buzzer. If you enter Display
Mode during an alarm condition, the buzzer will not be silenced, the LEDs will continue
to flash, and the vibrator will continue to pulse.
1 . From the
ALARM SETTING
menu, place the cursor next to
ALARM SILENCE
.
2 . Press and release POWER/ENTER. The Alarm Silence Option Screen appears.
3 . Use
▲
AIR or SHIFT
▼
(PANIC) to display the desired setting.
4 . Press and release POWER/ENTER to save the setting and return to the
ALARM
SETTING
menu.
Returning to the Maintenance Mode Menu (ESCAPE)
1 . From the
ALARM SETTING
screen, place the cursor next to
ESCAPE
.
2 . Press and release POWER/ENTER. The instrument will return to the main menu.
Turning the Pump On/Off Display Screen On and Off
(PUMP OFF DISP)
With
PUMP OFF DISP
set to
ON
, the Pump Off Screen appears in Display Mode and the
user can turn off the pump in order to conserve battery power.
With
PUMP OFF DISP
set to
OFF
(factory setting), the Pump Off Screen does not
appear in Display Mode and the user cannot turn the pump off.
1 . From the main menu, place the cursor next to
PUMP OFF DISP
.
2 . Press and release POWER/ENTER. The Pump Off Display Screen appears.