GX-6000 Operator’s Manual
Appendix A: Maintenance Mode • 133
ZERO FOLLOWER (pg.150)
For each channel:
• ON
• OFF
ON
ZERO SUPPRESS (pg.150)
0 - 3% of full scale
2% of full scale*
IR AUTO RANGE (pg.150)
• AUTO RANGE
• LOW ONLY
AUTO RANGE
CONFIRMATION (pg.150)
• ON
• OFF
OFF
CONFIRMATION INTERVAL (if
CONFIRMATION set to ON)
1 - 60 MIN
5 MIN
KEY TONE (pg.151)
• ON
• OFF
ON
INERT MODE (pg.151)
• ON
• OFF
OFF
L. /B. MODE (pg.152)
• OFF
• LEAK CHECK MODE
BAR HOLE MODE
• BAR HOLE MODE
• LEAK CHECK MODE
OFF
BAR HOLE TIME (pg.152)
• 30 SEC
• 45 SEC
• 60 SEC
30 SEC
CO DISPLAY (pg.152)
• ON
• OFF
OFF
MAN DOWN
(pg.153)
MAN DOWN (pg.153)
• ON
• OFF
OFF
PANIC (pg.153)
• ON
• OFF
ON
WARNING 1 TIME (pg.154)
10 - 75 SEC
60 SEC
WARNING 2 TIME (pg.154)
60 - 90 SEC
75 SEC
ALARM TIME (pg.154)
75 - 120 SEC
90 SEC
ESCAPE
return to main menu
N/A
Table 17: Maintenance Mode Menu Items
Menu Item (Page # of Description)
Choices or Action
Factory Setting