GX-6000 Operator’s Manual
Chapter 3: Operation • 29
NOTE:
If
BUMP DISP
is set to
OFF
, the bump test menu item will not appear in
Calibration Mode even though the instrument is prompting you to perform a
bump test. A bump test can always be performed in Maintenance Mode, if
necessary.
•
If the unit is due for bump testing and
BUMP EXPRD
is set to
CANNOT USE
,
then the following screen displays and the buzzer sounds in a double pulsing
pattern.
The GX-6000 cannot be used until a successful bump test has been performed.
Press and release the POWER/ENTER button to enter Calibration Mode. See
“Chapter 4: Calibration Mode” on page 69 for instructions to bump test the GX-
6000.
NOTE:
In this situation, even if the User password function has been turned on, no
password is required to perform a bump test.
NOTE:
If
BUMP DISP
is set to
OFF
, the bump test menu item will not appear in
Calibration Mode even though the instrument is prompting you to perform a
bump test. A bump test can always be performed in Maintenance Mode, if
necessary.
When you are done with the bump test and exit Calibration Mode, the unit will
begin the startup sequence. If the bump test was successful, the screen above will
not appear again until the unit is due for bump testing. If the bump test was not
successful, the screen above will again appear in the startup sequence.
4:07
BUMP DATE PAST
CANNOT USE
CAL MODE : ENTER