SLS Baseline 9000 FID  

Total Gas Detector Manual 

 
 
 

 

 

Summary of Contents for SLS Baseline 9000 FID

Page 1: ...SLS Baseline 9000 FID Total Gas Detector Manual Document Number GL SLS OEPS L4 94 Originators Lance Wheat Rev 0 Mar 2016 Approval No of Pages 38 SLS Baseline 9000 FID Total Gas Detector Manual ...

Page 2: ...1 of 36 SLS Baseline 9000 FID Total Gas Detector Manual Note The procedure in this manual should be performed in the order that the steps are written to properly operate the Baseline 9000 Total Gas Detector ...

Page 3: ... 1 3 Fuel 7 1 2 Baseline 9000 Baseline 8900 8 1 2 1 Air 8 1 2 2 Sample 9 1 2 3 Fuel 10 2 Calibration 11 2 1 Initial Flow Setup 11 2 2 Flow Calibration 12 2 2 1 Fuel 12 2 2 2 Air 15 2 2 3 Sample 18 2 3 Gas Calibration 21 3 WellWizard Connection 27 3 1 Hardware Setup 27 3 2 Tensor Communication 28 3 3 Unhiding Baseline 9000 Channels 31 4 Bill of Materials 35 ...

Page 4: ... Reading 14 Figure 14 Air Flow Zero Point 15 Figure 15 Air Flow Span Point 15 Figure 16 Air Flow Press Cnts 16 Figure 17 Air Flow Value 16 Figure 18 Air Flow Reading 17 Figure 19 Sample Flow Zero Point 18 Figure 20 Sample Flow Span Point 18 Figure 21 Sample Flow Press Cnts 19 Figure 22 Sample Flow Value 19 Figure 23 Sample Flow Reading 20 Figure 24 Zero Selection 21 Figure 25 Zero Value 21 Figure ...

Page 5: ...grams File 29 Figure 41 Enable Well Logging 30 Figure 42 Wits Diag 30 Figure 43 WWtext 31 Figure 44 Goto Channel 31 Figure 45 Mnemonic Entry 31 Figure 46 Channel Name 32 Figure 47 Unit Class 32 Figure 48 Hide Channel Status 32 Figure 49 Unhide Channel 33 Figure 50 Exit Channel 33 Figure 51 Channel Category 34 ...

Page 6: ...n are assuming the use of a Schmidlin Sonimix SX 3080 plus If not please verify the H2 volume generation capabilities before rig up to ensure the H2 generator being using has sufficient capacity to supply all equipment 1 1 Baseline 9000 Agilent 490 1 1 1 Air Figure 1 Baseline 9000 with Agilent 490 Air ...

Page 7: ...Page 6 of 36 SLS Baseline 9000 FID Total Gas Detector Manual 1 1 2 Sample Figure 2 Baseline 9000 with Agilent 490 Sample ...

Page 8: ...Page 7 of 36 SLS Baseline 9000 FID Total Gas Detector Manual 1 1 3 Fuel Figure 3 Baseline 9000 with Agilent 490 Fuel ...

Page 9: ...Page 8 of 36 SLS Baseline 9000 FID Total Gas Detector Manual 1 2 Baseline 9000 Baseline 8900 1 2 1 Air Figure 4 Baseline 9000 with Baseline 8900 Air ...

Page 10: ...Page 9 of 36 SLS Baseline 9000 FID Total Gas Detector Manual 1 2 2 Sample Figure 5 Baseline 9000 with Baseline 8900 Sample ...

Page 11: ...Page 10 of 36 SLS Baseline 9000 FID Total Gas Detector Manual 1 2 3 Fuel Figure 6 Baseline 9000 with Baseline 8900 Fuel ...

Page 12: ...pplied to the flow zone The Press Cnts shown on the screen refers to the pressure measurement from the sensor Figure 7 Baseline 9000 Rear Panel 2 1 Initial Flow Setup 1 From the Main screen press the or key Go to the Setup menu and select Flows 2 Press Enter to open the Setup Flow screen 3 To change a value use the key to erase the set value and the keypad to enter a new value Press Enter to selec...

Page 13: ...he Fuel Air and Sample input lines are disconnected 1 Connect a digital flow meter to the Detector Vent on the back of the TGD Figure 8 Detector Vent 2 From the Main screen press the or key Go to the Setup menu and select Flow Cal 3 Use the arrow to select Fuel Press Enter to open the Flow Calibration screen Zero Point Figure 9 Fuel Flow Zero Point 4 Allow time to stabilize minimum 2 minutes and t...

Page 14: ...Calibration screen is now shown Figure 10 Fuel Flow Span Point 6 Reconnect the Fuel line to the TGD and apply 20 psi to the fuel inlet port 7 Wait until the current pressure counts are equal to or close to the target pressure counts minimum 2 minutes Figure 11 Fuel Flow Press Cnts ...

Page 15: ...ange the target pressure counts Use the key to update the new value Wait for the pressure counts to stabilize at least 1 minute and re measure the flow at the detector vent Repeat until the measured flow is equal to the target flow 2 cc min Figure 12 Fuel Flow Value Figure 13 Fuel Flow Reading 9 When the measured flow equals the target flow press Enter to save the span point 10 If you make a mista...

Page 16: ...menu and select Flow Cal 3 Use the arrow to select Air Press Enter to open the Flow Calibration screen Zero Point Figure 14 Air Flow Zero Point 4 Allow time to stabilize minimum 2 minutes and then press Enter to accept the zero point NOTE Keep the detector elevated to ensure no liquid enters from the detector vent 5 The Span Point Calibration screen is now shown Figure 15 Air Flow Span Point 6 Rec...

Page 17: ...f the detector vent on the back panel and compare the measured value with the target flow shown on the display If necessary use the key and the numeric keypad to change the target pressure counts Use the key to update the new value Wait for the pressure counts to stabilize at least 1 minute and re measure the flow at the detector vent Repeat until the measured flow is equal to the target flow 2 cc...

Page 18: ...line 9000 FID Total Gas Detector Manual Figure 18 Air Flow Reading 9 When the measured flow equals the target flow press Enter to save the span point 10 If you make a mistake hit Escape to exit out without saving the changes ...

Page 19: ... Go to the Setup menu and select Flow Cal 3 Use the arrow to select Air Press Enter to open the Flow Calibration screen Zero Point Figure 19 Sample Flow Zero Point 4 Allow time to stabilize minimum 2 minutes and then press Enter to accept the zero point NOTE Keep the detector elevated to ensure no liquid enters from the detector vent 5 The Span Point Calibration screen is now shown Figure 20 Sampl...

Page 20: ...f the detector vent on the back panel and compare the measured value with the target flow shown on the display If necessary use the key and the numeric keypad to change the target pressure counts Use the key to update the new value Wait for the pressure counts to stabilize at least 1 minute and re measure the flow at the detector vent Repeat until the measured flow is equal to the target flow 2 cc...

Page 21: ...ne 9000 FID Total Gas Detector Manual Figure 23 Sample Flow Reading 9 When the measured flow equals the target flow press Enter to save the span point 10 If you make a mistake hit Escape to exit out without saving the changes ...

Page 22: ...own to Zero using the arrow keys and press the green enter key to select the Zero function Figure 24 Zero Selection 2 Use the left arrow to delete the previous value if the machine does not already indicate 0 00 percent and enter zero percent using the number pad Press the green enter key to save Figure 25 Zero Value ...

Page 23: ...rom the main menu under Calibration and press the green enter key to select it Figure 26 Span Selection 4 Set the span value to 10 00 using the number pad on the right of the machine and the left arrow to delete previous numbers Press the green enter key to save Figure 27 Span Value ...

Page 24: ...o the machine The zero may be set multiple times with each new selection overriding the previously selected zero value Once the zero value has become steady fluctuating between positive and negative values press the green enter key to save the calibration If a zero calibration is not required press the red key to skip this step and move onto the span calibration Figure 29 Zero Calibration ...

Page 25: ...alibration gas to the span port at 20 psi Figure 30 Span Connection 8 Use the left button to update the span value until the reading has settled likely fluctuating above and below the span value When the reading has settled press the green enter key to save the calibration This process may be skipped by pressing the red escape key if no span calibration is required Figure 31 Span Calibration ...

Page 26: ...apply 100 methane at least 100 cc min to the sample port using a vent valve to allow the machine to draw in the correct flow of gas Figure 32 Full Scale Test 10 If the reading is high i e 120 the air flow must be adjusted Go to setup on the main menu and scroll down to flows Figure 33 Flow Selection ...

Page 27: ...ease the value Press the green arrow then the red arrow two times until you are in the regular measuring display mode Figure 35 Air Flow Value 13 The reading should now be closer to 100 than the previous value Since the flow has been modified the 10 reading value might have been diminished as well so another span calibration is required Skip the zero step by pressing the red escape key after selec...

Page 28: ...he Baseline 9000 is reporting in static units by navigating the menu on the machine and selecting the options as shown Figure 36 Baseline 9000 Static Units 2 Ensure that the Baseline 9000 is using high precision by navigating the menu on the machine and selecting the options as shown Figure 37 Basline 9000 High Precision ...

Page 29: ...ID of the well UWID Figure 38 Query Well 3 Type cd ww well UWID config This will allow you to modify this well a UWID is the Unique Well ID for the particular well that is being set up The number may be copied by highlighting it from querying the well and left clicking then right clicking will paste it into the line Figure 39 Enter Well Directory 4 Type nano programs to modify the programs file of...

Page 30: ...oved from the beginning and the serial port number must be changed to the correct serial port for communication a In this line X is representative of which ever serial port number the Baseline unit is connected to Figure 40 Programs File 6 Press ctrl x to exit 7 Press y to save 8 Press enter to return from the programs file 9 Enter WWsetup 10 Go to Well Management ...

Page 31: ... SLS Baseline 9000 FID Total Gas Detector Manual 11 Enable Well logging for the well Figure 41 Enable Well Logging 12 Ensure that the device is transferring data by entering wits_diag p dev serX Figure 42 Wits Diag ...

Page 32: ... WellWizard version 7 1 Enter WWtext then using the arrow keys select the well and press ENTER Figure 43 WWtext 2 Press g for the Goto channel function Figure 44 Goto Channel 3 Enter BLTGM this is mnemonic for the corrected baseline total gas value Figure 45 Mnemonic Entry 4 When prompted for a new channel name press ENTER ...

Page 33: ...LS Baseline 9000 FID Total Gas Detector Manual Figure 46 Channel Name 5 Unit Class will remain the same press enter Figure 47 Unit Class 6 Using the arrow keys scroll over to Hide Channel Figure 48 Hide Channel Status ...

Page 34: ...f 36 SLS Baseline 9000 FID Total Gas Detector Manual 7 Press ENTER to change it from the default SET position Figure 49 Unhide Channel 8 Using the arrow keys go to the EXIT and press ENTER Figure 50 Exit Channel ...

Page 35: ...e 9000 FID Total Gas Detector Manual 9 Press Enter to confirm it is remaining under the same category Figure 51 Channel Category 10 Press Shift s to save the changes 11 Repeat for the following mnemonics o BL Total Gas Raw BLTG ...

Page 36: ...onized 5 gal 2182724316 Smooth Bore Seamless SS Tubing 1 8 OD 055 ID 6 feet 2182723PVC Polyurethane Tubing Clear 1 8 ID 1 4OD 25 feet 2182730Polyurethane Tubing Tinted Green 320in ID 1 2inOD 5 feet 2102563Air Compressor Assembly 1 each 2101500Compressor air 1HP 1725 RPM 2 5 GAL 60 dBA portable 1 each 2093737Adapter pipe 0 250 TBE x 0 250 male NPT 290 psi NP BRS 1 each 1754780Analyzer gas FID SER 9...