Chapter 9
9-28
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Figure 9-30: Rotary Encoder (9.5-in. GUI)
Rotary Encoder
(10.4-in. GUI’s)
Follow these steps to remove the rotary encoder (Figure 9-30). Reverse
removal steps to install.
NOTE:
Reinstall the rotary encoder according to the orientation shown.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the sensor PCB (see “Sensor PCB” on page 9-5).
3. Disconnect the rotary encoder harness from the J10 connector on the
MMI PCB (see “MMI PCB (10.4-in. GUI)” on page 9-18).
4. Remove the GUI assembly (see “GUI Assembly (Original 10.4-in.
5. Remove the ½-in. nut and star washer holding the rotary encoder to
the GUI assembly, then remove the encoder.
Front panel overlay
Rotary encoder cover
Rotary encoder