Chapter 8
8-10
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Performance Verification
2. Select Flow Rate (refer to Figure 8-6) and verify Auto-Triggering is
not
checked.
Figure 8-6: Trigger Options Screen
3. Use the up or down direction arrows to select 20/Lmin for Start Trigger
and 2/Lmin for End Trigger (refer to Figure 8-6).
4. When the appropriate parameters are selected, touch OK (refer to
Configuration Menu
Use this screen to save configurations and switch between different
configurations.
Saving Configurations
1. Touch the “Configuration” tab (refer to Figure 8-3).