Chapter 8
8-54
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Performance Verification
Test 4: Oxygen Flow Accuracy
Passed Value
Failed Value
Corrected
(after repair)
Value
Oxygen Position
display at 1 LPM (185-525
steps)
Steps
Steps
Steps
Oxygen Position
display at 180 LPM (refer
to altitude table)
Steps
Steps
Steps
Oxygen
at 5 LPM
Analyzer Flow
display (4.5-5.5 LPM)
LPM
LPM
LPM
Oxygen Flow
display (4.5-5.5 LPM)
LPM
LPM
LPM
Oxygen
at 10 LPM
Analyzer Flow
display (9-11 LPM)
LPM
LPM
LPM
Oxygen Flow
display (9-11 LPM)
LPM
LPM
LPM
Oxygen
at 20 LPM
Analyzer Flow
display (18-22 LPM)
LPM
LPM
LPM
Oxygen Flow
display (18-22 LPM)
LPM
LPM
LPM
Oxygen
at 50 LPM
Analyzer Flow
display (45-55 LPM)
LPM
LPM
LPM
Oxygen Flow
display (45-55 LPM)
LPM
LPM
LPM
Oxygen
at 100 LPM
Analyzer Flow
display (90-110 LPM)
LPM
LPM
LPM
Oxygen Flow
display (90-110 LPM)
LPM
LPM
LPM
Oxygen
at 120 LPM
Analyzer Flow
display (108-132 LPM)
LPM
LPM
LPM
Oxygen Flow
display (108-132 LPM)
LPM
LPM
LPM
Oxygen
at 165 LPM
Analyzer Flow
display (148.5-181.5 LPM)
LPM
LPM
LPM
Oxygen Flow
display (148.5-181.5 LPM)
LPM
LPM
LPM
Oxygen
at 0 LPM
Analyzer Flow
display (±0.1 LPM)
LPM
LPM
LPM
Oxygen Flow
display (±0.1 LPM)
LPM
LPM
LPM