Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-59
4. Connect cable O2 F/S to the oxygen flow sensor, then tie wrap it to the
circuit board. To avoid damaging the flow sensor, hand-tighten only
(do not tighten the tie wrap with a tool).
5. Install the enclosure brace.
6. Install the top enclosure (see “Top Enclosure” on page 9-4).
Inspiratory Manifold
Assembly
Follow these steps to remove the inspiratory manifold assembly (Figure 9-70,
Figure 9-71, Figure 9-72):
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove enclosure brace.
3. Remove the oxygen flow sensor (see “Oxygen Flow Sensor” on page 9-
57).
4. Remove the air flow sensor (see “Air Flow Sensor” on page 9-65).
5. Remove the vent port tubing from the inspiratory manifold.
6. Disconnect the tubes from the inspiratory, exhalation, and safety valve
solenoids.
7. Remove the green ground wire from the manifold.
8. Loosen the recessed 2.5-mm hex set screw on the outlet port, but do
not remove it.
When re-installing the outlet port, partially set the
setscrew into one of the indents. Turn the outlet port left and right to
be certain the setscrew is captured by the indent. Next, tighten the
setscrew all of the way and confirm the outlet port is tight.
9. Remove two 4-mm screws holding the inspiratory manifold to the
bottom enclosure.
10. Remove four 2.5-mm screws and washers surrounding the inspiratory
outlet port. Pull the outlet port from the inspiratory manifold.
Verify
check valve orientation at reinstallation.
11. Disconnect the cables labeled EXHAL SOL, INSP SOL, and SAFETY
SOL, and remove the inspiratory manifold.