Chapter 4
4-2
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Periodic Maintenance
Table 4-1: Schedule for Periodic Maintenance
At least every 250
hours
•
Air Inlet & Fan Filters
•
Inspect and clean. Some
environments cause a quicker
collection of lint and dust than others,
requiring maintenance more frequently
than every 250 hours.
Annually
•
Annual preventive
maintenance kit (P/N
1034840). Kit contents are
subject to change.
•
Install annual preventive maintenance
kit.
•
Clean ventilator interior and exterior
•
Complete performance verification
procedure
CAUTION: The annual preventive maintenance procedure is to be performed only
by a qualified service technician
12,500 hours
•
12,500 hour preventive
maintenance kit (P/N
1001733). Kit contents are
subject to change.
•
Install 12,500 hour preventive
maintenance kit
•
Clean ventilator interior and exterior
•
Complete performance verification
procedure
CAUTION: The 12,500 hour preventive maintenance procedure is to be
performed only by a qualified service technician.
As required
•
External oxygen sensor
•
Backup Battery
•
Replace and recalibrate new sensor by
running Extended Self-Test
• If you have the backup battery option,
refer to section 13 of the Operator’s
Manual for charging and maintenance
instructions