Chapter 9
9-4
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
To remove the oxygen water trap filter (Figure 9-6), unscrew the water trap,
then unscrew the filter element. Reverse to install a new filter.
Figure 9-6: Oxygen Water Trap Filter
Top Enclosure
Follow these steps to remove the top enclosure (Figure 9-7). Reverse removal
steps to install.
CAUTION:
Be careful not to pull or crimp any cables, tubes, or wires during removal
or installation.
1. Turn ventilator power OFF, then disconnect AC power and oxygen from
the ventilator.
2. Remove all accessories (including humidifiers, flex arms, etc.) and
external cables from the ventilator.
3. Remove six screws and washers (three on each side) from the side
rails. Two longer screws are closest to the front panel.
4. Remove the four screws and washers from the underside of the bottom
enclosure. Hang the enclosure off the work surface to remove the back
screws.
Oxygen water
trap filter