Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-63
replaced. For example, if replacing the crossover solenoid or oxygen
valve assembly, only remove the connector on the wire pair labeled
“CROSSOVER.”
Figure 9-73: Cutting Off Electrical Connector from Main PCB to Motor Cable
2. If necessary, affix the appropriate label from the cable extension kit to
identify the individual wire pair(s).
3. Insert the cut wire ends over both contacts in the crimp connector
(Figure 9-74) for each cut wire pair.
Figure 9-74: Installing Cut Wires into Crimp Connector
Cut close to the connector
Extension kit wires and connector
Crimp connector
Cut wires
Do not allow cut wires to protrude
outside crimp connector