Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-67
5. Disconnect the two MTR BV cables on the air valve from the mating
cable MTR BV.
6. Disconnect the silicone tube from the blower to the crossover solenoid
assembly.
7. Remove two 4-mm screws holding the air valve bracket, then carefully
remove the air valve.
Figure 9-78: Air Valve and Mesh Cup
Oxygen Valve
Assembly
Follow these steps to remove the oxygen valve assembly (Figure 9-79). Reverse
removal steps to install.
NOTE:
•
The oxygen regulator assembly must be installed before reinstalling
the oxygen valve assembly.
•
When reinstalling the oxygen valve, ensure that the oxygen regulator
and oxygen flow sensor are fully installed to the oxygen valve
assembly.
•
Always inspect o-rings between interconnecting parts and replace if
visibly damaged or rough to the touch. Always use new o-rings
supplied with replacement parts. Lightly lubricate o-rings with Krytox
Air valve
assembly
4-mm screws holding
air valve bracket
Mesh cup over air
valve inlet port