Chapter 9
9-96
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Downloading Ventilator Software
See Chapter 7 for complete software download instructions
1. Connect the service PC to the ventilator serial port using a null modem
cable.
2. Insert field upgrade CD-ROM and follow the onscreen instructions.
Use this command to prompt for the ventilator serial number (
x:
is the
CD-ROM drive, and a space precedes
-vs
):
(x:)\setup -vs
Calibrating the Screen
1. Cycle ventilator power while holding down the Exp. Hold and left
Options button.
2. The ventilator prompts you to calibrate the screen. Upon
acknowledgement, the software automatically detects and adjusts for
the 10.4-in. screen size.
3. After the ventilator confirms the screen calibration is complete, cycle
power to the ventilator and enter diagnostic mode.
4. Enter the software screen and confirm that the serial number is
correct.
5. Reset the altitude and compliance compensation settings in the user
configuration screen.
6. Perform and EST with the top enclosure removed. If the EST passes,
install the top enclosure (see “Top Enclosure” on page 9-4).
Enabling Options
Enable an option (see Chapter 7) only if the option upgrade is being installed
at this time.
Final Checkout
Complete the recommended tests in the performance verification procedure
(see Chapter 8).