Chapter 6
6-14
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Diagnostic Codes
3137
Air flow sensor (FS1) failure.
This code indicates that information
stored on FS1 cannot be read.
1. Enter diagnostic mode, then turn
ventilator off and on.
2. Replace CPU PCB.
3. Replace air flow sensor.
4. Replace sensor PCB.
5. Replace analog PCB.
6. Replace main PCB.
3138
Oxygen flow sensor (FS2) failure.
This code indicates that information
stored on FS2 cannot be read.
1. Enter diagnostic mode, then turn
ventilator off and on.
2. Replace CPU PCB.
3. Replace oxygen flow sensor.
4. Replace sensor PCB.
5. Replace analog PCB.
6. Replace main PCB.
3139
Exhalation flow sensor (FS3) failure.
This code indicates that information
stored on FS3 cannot be read.
1. Enter diagnostic mode, then turn
ventilator off and on.
2. Replace CPU PCB.
3. Replace exhalation flow sensor.
4. Replace sensor PCB.
5. Replace analog PCB.
6. Replace main PCB.
3140
Flow sensors cannot erase table.
This code indicates that flow sensor
calibration tables stored on the CPU
PCB cannot be erased.
Replace CPU PCB.
3141
Inhalation autozero solenoid cannot
open.
Block patient wye (test 1): patient
circuit pressurized to 50 cmH
2
O,
then inhalation autozero solenoid
commanded to energize (open), and
inspiration pressure transducer
(PT3) should measure 0 cmH
2
O.
This code indicates that PT3
measured outside the range from -3
to 3 cmH
2
O.
1. Check inhalation solenoid (Chapter
5).
2. Replace three-station solenoid
assembly.
3. Replace sensor PCB.
4. Replace analog PCB.
3142
Exhalation autozero solenoid cannot
open.
Block patient wye (test 1): patient
circuit pressurized to 50 cmH
2
O,
then exhalation autozero solenoid
commanded to energize (open), and
exhalation pressure transducer (PT2)
should measure 0 cmH
2
O. This code
indicates that PT2 measured outside
the range from -3 to 3 cmH
2
O.
1. Check inhalation solenoid (Chapter
5).
2. Replace three-station solenoid
assembly.
3. Replace sensor PCB.
4. Replace analog PCB.