Chapter 9
9-38
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Increased Minimum
Alarm Volume
Harness
Follow these steps to remove the increased minimum alarm volume harness
(Figure 9-42). Reverse removal steps to install as shown.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Disconnect the Alarm Volume Potentiometer from the Increased
Minimum Alarm Volume Harness.
3. Disconnect the Increased Minimum Alarm Volume Harness from J5 of
the MMI PCBA.
Figure 9-42: Increased Alarm Volume Harness
Backup Alarm
Follow these steps to remove the backup alarm (Figure 9-43). Reverse removal
steps to install.
NOTE:
Before reinstalling the backup alarm, inspect the EMI and mesh cooling
fan filters. If the filters are dirty, rinse with warm water and wipe or blow
dry.
Reinstall the EMI filter in the orientation shown.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the outer filter housing and mesh cooling inlet filter.
P/N 1022447