Chapter 8
Performance Verification
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
8-15
4. Touch
OK
to enter diagnostic mode.
5. Touch
EST
on the diagnostic screen (Figure 8-13).
Figure 8-13: Diagnostic Mode - Starting EST
6. Touch
Start EST
, then follow the prompts on the screen. EST must
pass before continuing.
NOTE:
If the EST fails a test, log the diagnostic code, measurement, and
tolerance parameters as displayed in the
Failure Data
window during EST
on the Performance Verification Data Form (section 8-7).
7. When EST is successfully completed, touch
User Config
to display the
User Configuration
screen (Figure 8-14).
Figure 8-14: Diagnostic Mode - User Configuration Screen
8. If necessary, touch
Compliance
to disable (gray background) the
compliance factor.
Touch
Start EST