Chapter 9
9-76
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Figure 9-86: Blower Assembly (Original and Updated)
Follow these steps to install the blower assembly (Figure 9-85, Figure 9-87):
NOTE:
The muffler assembly must be installed before reinstalling the blower
assembly.
1. Connect the transfer tube to the bottom of the cooling coil.
2. Place the blower assembly on the gasket so that blower outlet is
parallel to the enclosure wall. Fasten with four lock nuts (and L-
brackets, for the original version of the blower assembly).
NOTE:
The L-brackets (original blower assembly) should be perpendicular to the
blower after
tightening. Torque just until firm (do not overtighten).
3. Route the blower cables between the cooling coil and the enclosure
wall.
4. Connect the BLOWER J1 cable to connector J1 or J2 on the blower
motor controller PCB (Figure 9-83).
5. Connect the blue, yellow and orange wires to the green terminal block
connector J2 on the original blower motor controller PCB (Figure 9-
83), or to connector J1 on the updated blower motor controller PCB
(Figure 9-84).
6. Connect the thermostat wires to connectors J3 and J4 on the sensor
PCB (polarity is unimportant).
7. Connect the silicone tube from the top of the cooling coil to the
blower, then use the hose clamp to fasten.
Cooling coil
Blower
Assembly
L-brackets
and locknuts
Cooling coil
Blower
assembly
Locknuts
Original Blower Assembly
Updated Blower Assembly