Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-77
8. Connect the silicone tube to the brass barb connector on top of the
blower.
9. Place the shroud over the blower while pushing the transfer tube
through the hole, then install the grommet.
CAUTION:
Route the cables from the blower and fan through the cutout at the
bottom corner of the shroud.
Do not pinch wires under the blower shroud
.
10. Use the four 7-mm lock nuts to fasten the shroud. Connect the green
ground wire from the exhalation valve to the shroud stud. Torque just
until firm (do not overtighten).
11. Connect the small silicone tube from the blower to the crossover
solenoid assembly.
12. Install a new mesh cup on the top port of the air valve, then insert the
blower outlet hose over the mesh cup completely to the valve body.
13. Install enclosure brace.
14. Install top enclosure (see “Top Enclosure” on page 9-4).
Figure 9-87: Blower Assembly Position on Gasket
Blower Muffler
Assembly
Follow these steps to remove the blower muffler assembly (Figure 9-88):
1. Remove top enclosure (see “Top Enclosure” on page 9-4).
2. Remove enclosure brace.
3. Remove blower assembly (see “Blower Motor Controller PCB
4. Disconnect transfer tube from the bottom of cooling coil.
Blower outlet
Transfer tube
connects cooling
coil to air valve
inlet
Brass barb (silicone
tube to CV5 and
crossover solenoid)
Lock nuts
Gasket