Chapter 9
9-74
Esprit® Ventilator Service Manual © Respironics, Inc.
REF 580-1000-02 G
Component Removal/Installation
Blower Motor
Controller PCB
(Updated)
Follow these steps to remove the updated blower motor controller PCB from
the bottom enclosure (Figure 9-84). Reverse to install.
1. Remove top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the enclosure brace.
3. Remove four M3 nuts holding the blower motor controller PCB
bracket, then partially remove the bracket.
4. Disconnect the wires from the green terminal block (connector J2).
5. Disconnect the cables from the blower and sensor PCB harnesses.
6. Remove the entire bracketed blower motor controller PCB assembly.
Figure 9-84: Updated Blower Motor Controller PCB
Blower Assembly
Follow these steps to remove the blower assembly (Figure 9-85, Figure 9-86,
Figure 9-71).
1. Remove top enclosure (see “Top Enclosure” on page 9-4).
2. Remove enclosure brace.
3. Disconnect the blower outlet tube from the top port of the air valve.
Remove the mesh cup from inside the tube.
Connector J2
Red
Black
Orange
Yellow
Blue
To blower harness
To sensor PCB
harness