Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-39
3. Remove the four screws that hold the backup alarm bracket, then
separate components.
4. Disconnect the backup alarm from the J15 connector on the MMI
PCB.
5. Remove the collar from the alarm, then push the alarm from the
bracket.
Figure 9-43: Backup Alarm
AC Distribution Panel
WARNING:
The AC distribution panel contains high voltage components. To avoid injury,
verify that the AC and external DC power sources are disconnected from the
ventilator.
Follow these steps to remove the AC distribution panel (Figure 9-44, Figure 9-
45, Figure 9-46). Reverse removal steps to install.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the ten screws and washers that hold the AC distribution
panel to the enclosure. Note that one screw is longer (M3x12) and is
fastened on the inside of the enclosure with a 5.5-mm nut.
3. Remove the 10-mm nuts and washers holding the ground wires to the
ground lug and top enclosure, then remove the lug.
4. Remove the 2.5-mm screw holding the L-bracket, then lift the AC
distribution panel away from the top enclosure.