Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-75
4. Pry the grommet from the blower shroud.
5. Remove four 7-mm lock nuts holding the blower shroud to the bottom
enclosure.
6. Partially remove the shroud, pull the transfer tube into the shroud,
then completely remove the shroud.
7. Disconnect the silicone tube from the top of the cooling coil.
8. Remove four lock nuts (and L-brackets, for the original blower
assembly) holding the blower.
9. Loosen the hose clamp holding the silicone tube to blower outlet, then
disconnect the tube.
10. Disconnect the BLOWER cable from its connection on the blower
controller PCB.
11. Loosen the screws securing the orange, yellow, and blue wires from
the blower harness to the green terminal block connector J2 on the
blower motor controller PCB. Disconnect the wires (Figure 9-83,
Figure 9-84).
12. Disconnect the thermostat wires from connectors J3 and J4 on the
sensor PCB. Cut tie wraps and separate the harness back to the
blower.
13. Remove the blower.
Figure 9-85: Blower Shroud
Blower
shroud
Cooling
coil
Transfer
tube