Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-9
2. Remove the rotary encoder knob cover and loosen the 11-mm nut.
Remove the knob.
3. Remove the ½-in. nut and star washer.
4. Carefully separate the GUI from the enclosure. Disconnect the
interconnect harnesses and set the GUI face down on a soft, flat
surface.
5. Disconnect the harness from the backlight inverter PCB.
6. Remove the two Phillips head screws that hold the backlight inverter
PCB to the GUI.
Figure 9-11: Removing/Installing the Rotary Encoder Knob
Figure 9-12: Removing/Installing the Backlight Inverter PCB (10.4-in. GUI)
DC/DC Converter PCB
(Original 10.4-in.
GUI only)
Follow these steps to remove the DC/DC converter PCB (Figure 9-13) for 10.4-
in. GUI ventilators. Reverse removal steps to install.
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Disconnect the harness from the DC/DC converter PCB.
2.5-mm screws at bottom
corners of GUI
Knob cover
11-mm nut
Harnesses
Backlight
inverter PCB
10.4-in. LCD
assembly