Chapter 9
Component Removal/Installation
REF 580-1000-02 G
Esprit® Ventilator Service Manual © Respironics, Inc.
9-19
7. Remove the four screws holding the MMI PCB to the plate. Install
cables to the MMI PCB first, then insert the cables through the top
enclosure.
NOTE:
When reinstalling the ribbon cables to J6 and J11 (overlay LEDs), tug on
the cable to ensure that it is properly seated. When reinstalling the ribbon
cable to J12 and J13, ensure that no connector pins are visible, then
snap cover shut.
Figure 9-21: MMI PCB Connectors (10.4-in. GUI)
GUI Assembly
(9.5-in. GUI)
NOTE:
Because the original 9.5-in. GUI assembly is no longer available, the
10.4-in. GUI upgrade kit must be installed when replacing a VGA display,
touch frame, or keyboard.
Follow these steps to remove the 9.5-in. GUI assembly (Figure 9-22):
1. Remove the top enclosure (see “Top Enclosure” on page 9-4).
2. Remove the power supply shroud (see “Power Supply Fan/Shroud” on
Not
used
J11
J3
J12
J10
J1
J15
J5
J8
J4
J6
J13
J2
Not
used